MicroCraft

  • TDR Tester

Model Supporting High Frequency Probe

E1T6151 / E2T6151

E1T/E2T6151

Moving Probe TDR Measurement System
E1T6151 (1 probe) NEW
E2T6151 (2 probe) NEW
Impedance measurement system for high-end TDR users. It incorporates a high-precision mechanical stage and is capable of testing fine-pitch boards.

Detailed Specification

E1T6151 (1 Probe) E2T6151 (2 Probe)
Max. Test Area (vacuuming area)

17.7 x 16.1” (450 x 410 mm)

14.1 x 16.1” (360 x 410 mm)

Clamp System

Vacuum plate / TDR Multi Clamp

Vacuum plate / TDR Multi Clamp

Rotate Radius

325° (In case of Single End Probe)
235° (In case of Differential Probe set at 90°)
145° (In case of Differential Probe set at 180°)

325° (In case of Single End Probe)
235° (In case of Differential Probe set at 90°)
145° (In case of Differential Probe set at 180°)

Repeatable Accuracy X–Y

±0.16 mil (±4 µm)

±0.16 mil (±4 µm)

Repeatable Accuracy θ

Impedance Accuracy *1

50 Ω±1%

50 Ω±1%

Air Supply: Pressure

0.5 MPa

0.5 MPa

Air Supply: Volumetric flow rate

1.76 ft³/min. (50 ℓ/min.)

1.76 ft³/min. (50 ℓ/min.)

Power Supply
Single Phase AC200–240 V 50/60 Hz

6 A

8 A

Weight

2425 lbs. (1100 kg)

2535 lbs. (1150 kg)

Dimensions [ unit: inch(mm) ]

Required System

Control PC & Monitor *2

Operating Conditions 70–75°F (21–24°C), 30–70% relative humidity
* Do not use under the environment that has heavy condensing, dust, vibration, and/or corrosive gas.
Atmospheric Dew Point: −14°C or below (at 0.7 Mpa, dew point is 15°C or below)
Oil Mist Concentration: 1 mg/㎥ (ANR) or below

*1 Fluctuation at room temperature after calibration is ±2°C *2 Please contact our sales for details.

Supported Meter *3

Tektronix

DSA8300
DSA8200
TDS8000
11801B/C

Keysight Technologies

E5063A option TDR
E5071C+ENA option TDR
86100A/B/C
54750
N5224B

Teledyne LeCroy

WaveExpert100H+ST-20

MicroCraft

MZ30
MZ50

*3 Please contact our sales for details. Meters to be purchased separately.

TOP